Wearable smart sensor systems integrated on soft contact lenses for wireless ocular diagnostics

Author(s)
Kim, JooheeKim, MinjiLee, Mi-SunKim, KukjooJi, SangyoonKim, Yun-TaePark, JihunNa, KyungminBae, Kwi-HyunKim, Hong KyunBien, FranklinLee, Chang YoungPark, Jang-Ung
Issued Date
2017-04
DOI
10.1038/ncomms14997
URI
https://scholarworks.unist.ac.kr/handle/201301/21889
Fulltext
https://www.nature.com/articles/ncomms14997
Citation
NATURE COMMUNICATIONS, v.8, pp.14997
Abstract
Wearable contact lenses which can monitor physiological parameters have attracted substantial interests due to the capability of direct detection of biomarkers contained in body fluids. However, previously reported contact lens sensors can only monitor a single analyte at a time. Furthermore, such ocular contact lenses generally obstruct the field of vision of the subject. Here, we developed a multifunctional contact lens sensor that alleviates some of these limitations since it was developed on an actual ocular contact lens. It was also designed to monitor glucose within tears, as well as intraocular pressure using the resistance and capacitance of the electronic device. Furthermore, in-vivo and in-vitro tests using a live rabbit and bovine eyeball demonstrated its reliable operation. Our developed contact lens sensor can measure the glucose level in tear fluid and intraocular pressure simultaneously but yet independently based on different electrical responses.
Publisher
NATURE PUBLISHING GROUP
ISSN
2041-1723
Keyword
CARBON NANOTUBESCONDUCTING-POLYMERHYBRID STRUCTURESGLUCOSE SENSORTRANSPARENTGRAPHENEELECTRODESDESIGNIMMOBILIZATIONPERFORMANCE

qrcode

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.