Improvement of sensitivity and specificity for laminar BOLD fMRI with double spin-echo EPI in humans at 7 T

Author(s)
Han, SohyunEun, SeulgiCho, HyungjoonUludac, KamilKim, Seong-Gi
Issued Date
2021-11
DOI
10.1016/j.neuroimage.2021.118435
URI
https://scholarworks.unist.ac.kr/handle/201301/53366
Fulltext
https://www.sciencedirect.com/science/article/pii/S1053811921007102?via%3Dihub
Citation
NEUROIMAGE, v.241, pp.118435
Abstract
Mapping mesoscopic cortical functional units such as columns or laminae is increasingly pursued by ultra-high field (UHF) functional magnetic resonance imaging (fMRI). The most popular approach for high-resolution fMRI is currently gradient-echo (GE) blood oxygenation level-dependent (BOLD) fMRI. However, its spatial accuracy is reduced due to its sensitivity to draining vessels, including pial veins, whereas spin-echo (SE) BOLD signal is expected to have higher spatial accuracy, albeit with lower sensitivity than the GE-BOLD signal. Here, we introduce a new double spin-echo (dSE) echo-planar imaging (EPI) method to improve the sensitivity of SE-BOLD contrast by averaging two spin-echoes using three radiofrequency pulses. Human fMRI experiments were performed with slices perpendicular to the central sulcus between motor and sensory cortices at 7 T during fist-clenching with touching. First, we evaluated the feasibility of single-shot dSE-EPI for BOLD fMRI with 1.5 mm isotropic resolution and found that dSE-BOLD fMRI has higher signal-to-noise ratio (SNR), temporal SNR (tSNR), and higher functional sensitivity than conventional SE-BOLD fMRI. Second, to investigate the laminar specificity of dSE-BOLD fMRI, we implemented a multi-shot approach to achieve 0.8-mm isotropic resolution with sliding-window reconstruction. Unlike GE-BOLD fMRI, the cortical profile of dSE-BOLD fMRI peaked at similar to 1.0 mm from the surface of the primary motor and sensory cortices, demonstrating an improvement of laminar specificity in humans over GE-BOLD fMRI. The proposed multi-shot dSE-EPI method is viable for high spatial resolution UHF-fMRI studies in the pursuit of resolving mesoscopic functional units.
Publisher
Academic Press
ISSN
1053-8119
Keyword (Author)
High resolution fMRISpin-echo BOLDUltra-high fieldLayer specificity
Keyword
CEREBRAL BLOOD-VOLUMEOCULAR DOMINANCE COLUMNSGRADIENT-ECHOHUMAN BRAINHIGH-FIELDMAGNETIC-SUSCEPTIBILITYSPATIAL-RESOLUTIONFUNCTIONAL MRICONTRAST SENSITIVITYSIGNAL CHANGES