File Download

  • Find it @ UNIST can give you direct access to the published full text of this article. (UNISTARs only)
Related Researcher

김재준

Kim, Jae Joon
Circuits & Systems Design Lab.
Read More

Views & Downloads

Detailed Information

Cited time in webofscience Cited time in scopus
Metadata Downloads

A wireless ExG interface for patch-type ECG holter and EMG-controlled robot hand

Author(s)
Lee, KwangmukChoi, Yun YoungKim, Dae JungChae, Hee YoungPark, KyeonghwanOh, Young MinWoo, Sung HoonKim, Jae Joon
Issued Date
2017-08
DOI
10.3390/s17081888
URI
https://scholarworks.unist.ac.kr/handle/201301/22471
Fulltext
http://www.mdpi.com/1424-8220/17/8/1888
Citation
SENSORS, v.17, no.8, pp.1888
Abstract
This paper presents a wearable electrophysiological interface with enhanced immunity to motion artifacts. Anti-artifact schemes, including a patch-type modular structure and real-time automatic level adjustment, are proposed and verified in two wireless system prototypes of a patch-type electrocardiogram (ECG) module and an electromyogram (EMG)-based robot-hand controller. Their common ExG readout integrated circuit (ROIC), which is reconfigurable for multiple physiological interfaces, is designed and fabricated in a 0.18 μm CMOS process. Moreover, analog pre-processing structures based on envelope detection are integrated with one another to mitigate signal processing burdens in the digital domain effectively
Publisher
MDPI AG
ISSN
1424-8220
Keyword (Author)
motion artifactreadout integrated circuitECG holterrobot-hand controllerenvelope detectorlevel detector
Keyword
PROCESSORSYSTEM

qrcode

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.