Optimization of sparse phase encodings for variable repetition-delay turbo-spin echo (TSE) T1 measurements for preclinical applications

Author(s)
Lee, DongKyuHan, SohyunCho, Hyungjoon
Issued Date
2017-01
DOI
10.1016/j.jmr.2016.11.004
URI
https://scholarworks.unist.ac.kr/handle/201301/20742
Fulltext
http://www.sciencedirect.com/science/article/pii/S1090780716302300
Citation
JOURNAL OF MAGNETIC RESONANCE, v.274, pp.57 - 64
Abstract
A variable repetition-delay (TR) spin echo sequence with repeated refocusing pulses, i.e., a variable TR turbo-spin echo (TSE), provides an attractive means of acquiring an accurate T1 map information that is free from gradient echo based artifacts such as magnetic field inhomogeneities particularly for ultra-high field (at 7T and above) preclinical applications. However, the applicability of multi-slice TSE sequences is often limited by signal distortion from T2 relaxation due to echo-train acquisitions for short T2 tissues, inter-slice cross talks and magnetization transfer (MT) from repetitive slice-selective 180° pulse, and extended scan times with multiple TR excitations. These TSE shortcomings are difficult to remedy for preclinical applications, where small sizes of target organs usually limit the slice-gap control with restricted parallel imaging capabilities. In this study, compressed-sensing-assisted turbo-spin echo (CS-TSE) acquisitions for variable TR T1 measurements at 7T preclinical scanner were implemented to reduce the echo-trains by sparse phase encodings. Following the sparse signal simulation and sampling scheme optimization, the measured T1 values from CS-TSE and TSE were compared for phantoms, ex vivo, and in vivo subjects. The phantom T1 values from CS-TSE and TSE were identical to those from the inversion recovery spin echo. For both ex vivo and in vivo multi-slice T1 mapping, the shortened echo-trains of CS-TSE relieved the T2 relaxation, reduced the inter-slice interferences of multi-slice acquisition, and made room for additional slice encodings while maintaining a shorter scan time than the conventional TSE at the expense of local image smoothness from CS regularizations.
Publisher
ACADEMIC PRESS INC ELSEVIER SCIENCE
ISSN
1090-7807
Keyword (Author)
Compressed sensingTurbo-spin echoT-1 measurementsTSECS-TSERARE
Keyword
MRT1RECONSTRUCTIONRELAXATIONACCURACYCONTRAST